%0 Conference Paper %A Nguyen, Xuan Toan %A Do, Thi Huong Giang %A Nguyen, Huu Duc %B The 3rd International Conference on Advanced Materials and Nanotechnology (ICAMN) %C Hanoi, Vietnam %D 2016 %F SisLab:2269 %P 376 %T Measurement of Iron-Oxide Fluids Using a Magnetoelectric Sensor %U https://eprints.uet.vnu.edu.vn/eprints/id/eprint/2269/ %X Iron oxide fluids have been measured by using a magnetoelectric (ME) effect - based magnetic sensor with a long-type sandwiched composite of Metglass/PZT/Meglass laminates of 1×15 mm2 in dimension. In homogeneous magnetic fields, the sensor exhibits a sensitivity of 307.453 mV/Oe. The minimum measurable signal (i.e. noise level) of this experimental setup was found as small as 0.782 mV, which reveals a possible detection limit of 3×10−8 emu . Our report presents the simulation and the measurement of magnetic fluids of various magnetic Fe3O4 nanoparticle concentrations. The results open a possibility of using this sensor for medical and bioengineering applications.